|
Setup issues:
There
are numerous logging parameters that can be changed. The settings
are saved in the
instrument memory. We usually setup the instrument directly through
a terminal program
but hydrosoft can be used as well. Below are some tips to avoid
common mistakes:
- The AGAIN
(auto gain) function must be on (1) during a deployment. The
calibration
procedure requires the AGAIN function to be switched off.
So there is a risk that the
autogain be accidentally left off after the calibration.
- When
running the instrument on external power through a data logger
(like Modaps or HydroDAS):
1- Make sure you set the SleepOnFull function to 0. If on
1 the instrument will stop
logging data when its internal memory gets full, even if the
data is actually collected
externally. The internal memory can gets full on a very long
cast or if old casts still in
the memory were not deleted prior to the deployment. The SleepOnFull
function can
be set though the LOG command.
2- Make sure the AUTOSTART function is on 1. The instrument
will then start logging
as soon as external power it switch on.
3- If changing
the the LOG parameters or turning on/off the florescence channels
make
sure you make these changes nonvolatile with the STORELOG
and/or STOREFL
command.
Data
Issues:
The data obtained
from the 676 channel should be used with caution. 676 nm light
stimulates fluorescence. Because of large optical filter bandwidth
(~20nm FWHM), part of the scattering picked up by the 676 channel
comes actually from stimulated fluorescence emission at a slightly
longer wavelength. The fluorescence contribution to the backscattering
signal can be significant.
|